Monte Carlo simulation of Cherenkov imaging for Total Skin Electron Treatment with CT DICOM realistic patient geometry.
Monte Carlo simulation of Cherenkov imaging for Total Skin Electron Treatment with CT DICOM realistic patient geometry.
复制标题
使用 CT DICOM 真实患者几何形状进行全皮肤电子治疗切伦科夫成像的蒙特卡罗模拟。
DOI:
10.1117/12.2609027
复制
发表时间:
2022
期刊:
影响因子:
--
通讯作者:
Zhu,TimothyC
中科院分区:
文献类型:
--
作者:
Zhong,Weili;Ong,Yihong;Miao,Tianshu;Pogue,BrianW;Zhu,TimothyC
This Monte Carlo (MC) simulation study provides an evaluation of dose uniformity in a patient and the difference between dose and Cherenkov distributions, which is invaluable in developing conversion factors to relate observed Cherenkov images to actual dose distributions for TSET patients. This MC simulations with TOPAS is performed using realistic patient geometries obtained with a 3D scanner during total skin electron treatments (TSET) at UPenn. For each treatment posture in the Stanford technique, the differences between Cherenkov photon distributions and dose distributions produced in MC are consistent with the differences observed between a Cherenkov imaging camera and in-vivo dose measurement with OSLD on patient skin. According to MC studies of a flat rectangular PVC board, the difference between Cherenkov and dose is mostly due to the spoiler. This is confirmed by observing consistent dose and Cherenkov distributions in clinical measurements on a PVC board without the spoiler. The accumulated dose and Cherenkov distributions for each patient are obtained by projecting the MC output of the 6 postures of the TSET treatment together onto a finite element model of the patient.