The noise of membrane capacitance measurements in the whole-cell recording configuration
The noise of membrane capacitance measurements in the whole-cell recording configuration
复制标题
DOI:
10.1016/s0006-3495(00)76464-2
复制
发表时间:
2000-10-01
影响因子:
3.4
通讯作者:
Gillis, KD
中科院分区:
文献类型:
--
作者:
Chen, P;Gillis, KD
High-resolution measurement of membrane capacitance in the whole-cell-recording configuration can be used to detect small changes in membrane surface area that accompany exocytosis and endocytosis, We have investigated the noise of membrane capacitance measurements to determine the fundamental limits of resolution in actual cells in the whole-cell mode. Two previously overlooked sources of noise are particularly evident at low frequencies. The first noise source is accompanied by a correlation between capacitance estimates, whereas the second noise source is due to "1/f-like" current noise. An analytic expression that summarizes the noise from thermal and 1/f sources is derived, which agrees with experimental measurements from actual cells over a large frequency range. Our results demonstrate that the optimal frequencies for capacitance measurements are higher than previously believed. Finally, we demonstrate that the capacitance noise at high frequencies can be reduced by compensating for the voltage drop of the sine wave across the series resistance.