Purpose: To implement a short-duration magnetization preparation sequence, which consists of a saturation followed by multiple inversion pulses, for imaging of short-T-1 species and suppression of long-T-1 species.Materials and Methods: Computer optimizations were performed to derive preparation schemes that 1) suppress long-T-1 background species with T-1 >= 250 msee, 2) maximize the M-z of contrast-enhanced (CE) structures with T-1 250 msec, and about a 30% reduction for 20 msec < T-1 < 50 msec. The phantom studies validated the performance of the optimized sequences. Coronary artery angiograms (380 msec for preparation and image acquisition) demonstrated signal-to-noise ratios (SNRs) in the range of 13-15.5 and contrast-to-noise ratios (CNRs) in the range of 6.3-7.1 in the CE coronary vessels.Conclusion: This work demonstrates that fast magnetization-driven preparation schemes can be implemented for fast imaging of CE coronary vessels with efficient saturation of background species.