Real-time in vivo photoacoustic and ultrasound imaging

Real-time in vivo photoacoustic and ultrasound imaging
复制标题

DOI:
10.1117/1.3005421
复制
发表时间:
2008-09-01
影响因子:
3.5
通讯作者:
van Leeuwen, Ton G.
van Leeuwen, Ton G.
中科院分区:
医学3区
文献类型:
--
作者:
Kolkman, Roy G. M.;Brands, Peter J.;van Leeuwen, Ton G.

文献摘要

被引文献

相似文献

设计并搭建了实时光声成像系统。该系统基于商业上可用的超声成像系统。它可以达到8帧/秒的帧速率。志愿者手中的血管系统被成像,所产生的光声图像与超声图像相结合。通过对手部皮下血管分支的成像,展示了具有混合超声探头的实时光声成像系统。(C)2008年光学仪器工程师学会。[DOI:10.1117/1.3005421]
A real-time photoacoustic imaging system is designed and built. This system is based on a commercially available ultrasound imaging system. It can achieve a frame rate of 8 frames/sec. Vasculature in the hand of a human volunteer is imaged, and the resulting photoacoustic image is combined with the ultrasound image. The real-time photo acoustic imaging system with a hybrid ultrasound probe is demonstrated by imaging the branching of subcutaneous blood vessels in the hand. (C) 2008 Society of Photo-Optical Instrumentation Engineers. [DOI: 10.1117/1.3005421]