Comparison of fat quantification methods: A phantom study at 3.0T

Comparison of fat quantification methods: A phantom study at 3.0T
复制标题

DOI:
10.1002/jmri.21201
复制
发表时间:
2008-01-01
影响因子:
4.4
通讯作者:
Langton, Christian M.
Langton, Christian M.
中科院分区:
医学2区
文献类型:
--
作者:
Bernard, Clare P.;Liney, Gary P.;Langton, Christian M.

文献摘要

被引文献

相似文献

目的:比较不同成像方法在3.0T条件下的单体素磁共振波谱(MRS)对体模脂肪含量的定量分析。材料和方法:采用GE Signa系统进行成像和波谱分析。构建了11个新型均质脂肪水体模,脂肪含量从0%到100%不等。这些使用三种技术成像,并与单体素非水抑制MRS。像素的脂肪分数的像素地图进行比较,并与MRS测定的测量值的平均值。结果:所有成像方法与MRS的相关性均显著(P < 0.001):IDEAL(r(2)= 0.985)、IOP(r(2)= 0.888)、WS(r(2)= 0.939)、FS(r(2)= 0.938)。此外,IDEAL提供了脂肪分数的无伪影图,具有上级均匀性。在体内使用IDEAL的结果产生的值之间的9%至16%的相应的MRS value.Conclusion:这项工作表明,成像可以利用作为一个高分辨率的替代MRS的脂肪含量的定量。将来,我们打算在涉及脂肪测量的临床研究中用IDEAL代替MRS。
Purpose: To compare different imaging methods with single-voxel MR spectroscopy (MRS) for the quantification of fat content in phantoms at 3.0T.Materials and Methods: Imaging and spectroscopy was performed on a GE Signa system. Eleven novel homogeneous fat-water phantoms were constructed with variation in fat content from 0% to 100%. These were imaged using three techniques and compared with single-voxel non-water-suppressed MRS. Pixel-by-pixel maps of fat fraction were produced and mean values compared to MRS-deter mined measurements. Preliminary in vivo examinations were subsequently performed in the breast and spine to compare the best imaging technique with MRS.Results: All imaging methods significantly correlated with MRS (P < 0.001): IDEAL (r(2) = 0.985), IOP (r(2) = 0.888), WS (r(2) = 0.939), and FS (r(2) = 0.938). In addition, IDEAL provided artifact-free maps of fat fraction with superior uniformity. In vivo results using IDEAL produced values that were between 9% to 16% of the corresponding MRS values.Conclusion: This work demonstrates that imaging may be utilized as a high-resolution alternative to MRS for the quantification of fat content. In the future we intend to replace MRS with IDEAL in our clinical studies involving fat measurement.