Diagnostic Accuracy of Angiography-Based Quantitative Flow Ratio Measurements for Online Assessment of Coronary Stenosis
Diagnostic Accuracy of Angiography-Based Quantitative Flow Ratio Measurements for Online Assessment of Coronary Stenosis
复制标题
基于血管造影的定量流量比测量用于冠状动脉狭窄在线评估的诊断准确性
DOI:
10.1016/j.jacc.2017.10.035
复制
发表时间:
2017-12-26
影响因子:
24
通讯作者:
Hu, Shengshou
中科院分区:
文献类型:
--
作者:
Xu, Bo;Tu, Shengxian;Hu, Shengshou
BACKGROUND Quantitative flow ratio (QFR) is a novel angiography-based method for deriving fractional flow reserve (FFR) without pressure wire or induction of hyperemia. The accuracy of QFR when assessed online in the catheterization laboratory has not been adequately examined to date.OBJECTIVES The goal of this study was to assess the diagnostic performance of QFR for the diagnosis of hemodynamically significant coronary stenosis defined by FFR = 2 mm according to visual estimation. QFR, quantitative coronary angiography (QCA), and wire-based FFR were assessed online in blinded fashion during coronary angiography and re-analyzed offline at an independent core laboratory. The primary endpoint was that QFR would improve the diagnostic accuracy of coronary angiography such that the lower boundary of the 2-sided 95% confidence interval (CI) of this estimate exceeded 75%.RESULTS Between June and July 2017, a total of 308patients were consecutively enrolled at 5 centers. Online QFR and FFR results were both obtained in 328 of 332 interrogated vessels. Patient-and vessel-level diagnostic accuracy of QFR was 92.4%(95% CI: 88.9% to 95.1%) and 92.7%(95% CI: 89.3% to 95.3%), respectively, both of which were significantly higher than the pre-specified target value (p