Performance of human body communication-based wearable ECG with capacitive coupling electrodes
Performance of human body communication-based wearable ECG with capacitive coupling electrodes
复制标题
DOI:
10.1049/htl.2016.0023
复制
发表时间:
2016-09-01
影响因子:
2.1
通讯作者:
Wang, Jianqing
中科院分区:
文献类型:
--
作者:
Sakuma, Jun;Anzai, Daisuke;Wang, Jianqing
Wearable electrocardiogram (ECG) is attracting much attention in daily healthcare applications, and human body communication (HBC) technology provides an evident advantage in making the sensing electrodes of ECG also working for transmission through the human body. In view of actual usage in daily life, however, non-contact electrodes to the human body are desirable. In this Letter, the authors discussed the ECG circuit structure in the HBC-based wearable ECG for removing the common mode noise when employing non-contact capacitive coupling electrodes. Through the comparison of experimental results, they have shown that the authors' proposed circuit structure with the third electrode directly connected to signal ground can provide an effect on common mode noise reduction similar to the usual drive-right-leg circuit, and a sufficiently good acquisition performance of ECG signals.