Correction of off resonance-related distortion in echo-planar imaging using EPI-based field maps

Correction of off resonance-related distortion in echo-planar imaging using EPI-based field maps
复制标题

DOI:
10.1002/mrm.1910390223
复制
发表时间:
1998-02-01
影响因子:
3.3
通讯作者:
Frank, LR
Frank, LR
中科院分区:
医学3区
文献类型:
--
作者:
Reber, PJ;Wong, EC;Frank, LR

文献摘要

被引文献

相似文献

We present, here, a simple method for measurement and correction of off-resonance related geometric distortion in echo-planar imaging (EPI), This method uses high signal-to-noise ratio (SNR) EPI-based field maps, rapidly acquired using a series of gradient recalled images collected across a range of TE values, This field map is distorted in the same manner as the EPI images to be unwarped, providing a direct look-up table for the correct location of each pixel of data, This method adds very little scan time and is robust and easy to implement.