Validation of Physical Activity Monitors in Individuals with Diabetes: Energy Expenditure Estimation by the Multisensor SenseWear Armband Pro3 and the Step Counter Omron HJ-720 Against Indirect Calorimetry During Walking

Validation of Physical Activity Monitors in Individuals with Diabetes: Energy Expenditure Estimation by the Multisensor SenseWear Armband Pro3 and the Step Counter Omron HJ-720 Against Indirect Calorimetry During Walking
复制标题

DOI:
10.1089/dia.2012.0235
复制
发表时间:
2013-05-01
影响因子:
5.4
通讯作者:
Slaby, Krystof
Slaby, Krystof
中科院分区:
医学3区
文献类型:
--
作者:
Machac, Stanislav;Prochazka, Michal;Slaby, Krystof

文献摘要

被引文献

相似文献

背景资料:本研究的目的是测试SenseWear(R)臂带Pro 3(SWA)的能量消耗估计值与(BodyMedia,Pittsburgh,PA)和Omron HJ-720(Omron Healthcare,京都,日本)步计数器与间接测热法(IC)作为1型和2型糖尿病老年人在跑步机上行走时的金标准。总共有6名男性(60.3 ± 3.1岁)和13名女性(51.1 ± 11.0岁)患有1型或2型糖尿病。每名受试者在跑步机上进行三次15分钟的步行,速度和坡度的不同组合(3 km/h,0%; 4 km/h,0%; 5 km/h,5%)。能量消耗(EE)同时测量的SWA,欧姆龙,和IC。使用平均高估/低估和Pearson相关系数对测试方法和IC之间的一致性进行统计学评价。在3 km/h的速度和0%的坡度下,发现SWA的平均高估为+81.19 +/-23.81欧姆龙为+70.51 +/-20.91%(r = 0.77,P < 0.001)。在4km/h速度和0%坡度下,SWA的平均高估率为+78.18 ± 33.96%(r = 0.63,P < 0.01),Omron的平均高估率为+75.77 ± 33.36%(r = 0.52,P < 0.05)。在5 km/h速度和5%坡度的高强度运动水平下,SWA的平均低估率为-7.88 ± 16.86(r = 0.74,P < 0.001),欧姆龙为-7.37 +/- 16.07%结论:两种方法均明显高估了水平步行时的EE值,而低估了快速上坡步行时的EE值。
Background: The purpose of this study is to test the agreement between energy expenditure estimate of the SenseWear (R) Armband Pro3 (SWA) (BodyMedia, Pittsburgh, PA) and the Omron HJ-720 (Omron Healthcare, Kyoto, Japan) step counter with indirect calorimetry (IC) as a gold standard in older individuals with type 1 and type 2 diabetes mellitus while walking on a treadmill.Subjects and Methods: In total, six men (60.3 +/- 3.1 years old) and 13 women (51.1 +/- 11.0 years old) with type 1 or type 2 diabetes mellitus were included in the study. Each subject performed three 15-min walking sessions with different combinations of speed and incline (3 km/h, 0%; 4 km/h, 0%; 5 km/h, 5%) on a treadmill. Energy expenditure (EE) was simultaneously measured by the SWA, Omron, and IC. Mean over-/underestimation and Pearson's correlation coefficients were used for statistical evaluation of the agreement between tested methods and IC.Results: At the speed of 3 km/h with 0% incline, mean overestimation of +81.19 +/- 23.81% was found for SWA (r = 0.79, P < 0.001) and +70.51 +/- 20.91% for Omron (r = 0.77, P < 0.001). At the speed of 4 km/h and 0% incline, mean overestimation found for SWA was +78.18 +/- 33.96% (r = 0.63, P < 0.01) and +75.77 +/- 33.36% for Omron (r = 0.52, P < 0.05). At the level of high-intensity exercise at the speed of 5 km/h and 5% incline, mean underestimation was -7.88 +/- 16.86% for SWA (r = 0.74, P < 0.001) and -7.37 +/- 16.07% for Omron (r = 0.75, P < 0.001).Conclusions: Both methods led to considerable overestimation of calculated EE in level walking and a relatively minor underestimation during fast uphill walking.