Integrated intravascular ultrasound and photoacoustic imaging scan head
Integrated intravascular ultrasound and photoacoustic imaging scan head
复制标题
DOI:
10.1364/ol.35.002892
复制
发表时间:
2010-09-01
期刊:
影响因子:
3.6
通讯作者:
Li, Pai-Chi
中科院分区:
文献类型:
--
作者:
Hsieh, Bao-Yu;Chen, Sung-Liang;Li, Pai-Chi
The combination of intravascular ultrasound and intravascular photoacoustic imaging has been proposed for improving the diagnosis of arterial diseases. We describe a novel scan-head design for implementing such multimodality imaging. The proposed device has the potential to achieve a sufficiently small size for clinical intravascular applications. The design aims for efficient image data acquisition for facilitating real-time three-dimensional imaging and reducing the required laser pulse repetition frequency. The integrated scan head consists of a single-element, ring-shaped transducer for sideward ultrasound transmission, a multimode fiber with a cone-shaped mirror for optical illumination, and a single polymer microring with mechanical scanning. The phantom imaging and some experimental results are presented. A microring array can be realized in the future to achieve high-frame-rate intravascular multimodality imaging. (C) 2010 Optical Society of America