A simple high-sensitivity technique for purity analysis of xenon gas
A simple high-sensitivity technique for purity analysis of xenon gas
复制标题
一种简单的高灵敏度氙气纯度分析技术
DOI:
10.1016/j.nima.2010.04.152
复制
发表时间:
2010
影响因子:
1.4
通讯作者:
Y. Yen
中科院分区:
文献类型:
--
作者:
D. Leonard;D. S. Leonard;A. Dobi;C. Hall;L. Kaufman;T. Langford;S. Slutsky;Y. Yen
We report on the development and performance of a high-sensitivity purity-analysis technique for gaseous xenon. The gas is sampled at macroscopic pressure from the system of interest using a UHV leak valve. The xenon present in the sample is removed with a liquid-nitrogen cold trap, and the remaining impurities are observed with a standard vacuum mass-spectroscopy device. Using calibrated samples of xenon gas spiked with known levels of impurities, we find that the minimum detectable levels of N2, O2, and methane are 1×10−9, 160×10−12, and 60 ×10−12g/g, respectively. This represents an improvement of about a factor of 10000 compared to measurements performed without a cold trap.