A simple high-sensitivity technique for purity analysis of xenon gas

A simple high-sensitivity technique for purity analysis of xenon gas
复制标题

一种简单的高灵敏度氙气纯度分析技术

DOI:
10.1016/j.nima.2010.04.152
复制
发表时间:
2010
影响因子:
1.4
通讯作者:
Y. Yen
Y. Yen
中科院分区:
物理与天体物理3区
文献类型:
--
作者:
D. Leonard;D. S. Leonard;A. Dobi;C. Hall;L. Kaufman;T. Langford;S. Slutsky;Y. Yen

文献摘要

被引文献

相似文献

我们报告的发展和性能的高灵敏度纯度分析技术的气态氙。使用UHV泄漏阀在宏观压力下从感兴趣的系统中对气体进行采样。用液氮冷阱除去样品中存在的氙,用标准真空质谱仪观察剩余的杂质。使用已知杂质水平的氙气校准样品,我们发现N2,O2和甲烷的最小可检测水平分别为1×10−9,160×10−12和60 ×10− 12 g/g。这表示与在没有冷阱的情况下进行的测量相比改进了约10000倍。
We report on the development and performance of a high-sensitivity purity-analysis technique for gaseous xenon. The gas is sampled at macroscopic pressure from the system of interest using a UHV leak valve. The xenon present in the sample is removed with a liquid-nitrogen cold trap, and the remaining impurities are observed with a standard vacuum mass-spectroscopy device. Using calibrated samples of xenon gas spiked with known levels of impurities, we find that the minimum detectable levels of N2, O2, and methane are 1×10−9, 160×10−12, and 60 ×10−12g/g, respectively. This represents an improvement of about a factor of 10000 compared to measurements performed without a cold trap.