Direct, simplified, and sensitive assay of angiotensin II in plasma extracts performed with a high-affinity monoclonal antibody.
Direct, simplified, and sensitive assay of angiotensin II in plasma extracts performed with a high-affinity monoclonal antibody.
复制标题
使用高亲和力单克隆抗体对血浆提取物中的血管紧张素 II 进行直接、简化且灵敏的测定。
DOI:
10.1093/clinchem/38.10.1963
复制
发表时间:
1992
影响因子:
9.3
通讯作者:
P. Corvol
中科院分区:
文献类型:
--
作者:
D. Simon;B. Romestand;H. Huang;G. Badouaille;J. Fehrentz;B. Pau;J. Marchand;P. Corvol
A very simple, fast, and sensitive RIA of angiotensin (Ang) II has been developed, based on a monoclonal antibody with high affinity and specificity, making possible the direct measurement of circulating Ang II in human plasma after solid-phase extraction. The purified monoclonal antibody 4D8 has an association constant of 1.3 x 10(11) L/mol with Ang II and a cross-reactivity of < 1% for Ang I. The assay can detect as little as 0.8 fmol of Ang II in 2 mL of plasma and is not influenced by the presence of Ang I. Analytical recoveries between 112% and 116% were obtained for Ang II added to human plasma at physiological concentrations. Comparison of the RIA with a reversed-phase, high-performance liquid chromatographic method followed by RIA to measure Ang II in human plasma samples from normal and hypertensive subjects--and from normotensive subjects before and after an acute inhibition of angiotensin-converting enzyme with captopril (50 mg)--showed a high degree of correlation (r2 = 0.93) between the two methods.