A calibration standard for laser‐Doppler perfusion measurements
A calibration standard for laser‐Doppler perfusion measurements
复制标题
激光多普勒灌注测量的校准标准
DOI:
--
复制
发表时间:
1995
期刊:
影响因子:
--
通讯作者:
R. Maniewski
中科院分区:
文献类型:
--
作者:
A. Liebert;M. Leahy;R. Maniewski
The objective of the present study was the development of a new calibration standard for the laser‐Doppler perfusion monitors. The beat frequency spectrum obtained from a solution of latex spheres undergoing Brownian motion was compared to that obtained from a porous polyethylene material containing the same solution. The mixture of static and dynamic scattering events in the latter case was expected to produce a spectrum similar to that obtained in the real human tissue. The results obtained demonstrate clearly the difference between the signals recorded from the commonly used motility standard which are the result of multiple Doppler shifting and homodyne mixing, and those generated from our enhanced model which predominantly result from single Doppler shifting and heterodyne mixing.