背景资料:方法:在Sysmex XE-5000全自动生化分析仪上进行红细胞(RBC)和白色细胞(WBC)的体液(BF)模式检测。BF(差异)计数(139例患者样本和87例正常样本)进行测量,并与Fuchs-Rosenthal室和染色的细胞离心涂片进行比较。注意到使用BF模式进行的扩展细胞计数的WBC检测限(CV 20%)提高至10 × 10(6)/L。对于大多数BF,观察到与手动方法的高度一致性[脑脊液(CSF)的平均偏倚为+2至6x 10(6)/L,其他液体的平均偏倚为-1至12 x10(6)/L]。在CSF中,与手动方法相比,BF模式计数更多的WBC(多态性核细胞)(平均偏倚+5至6x 10(6)/L),尤其是在细胞计数较低的样本中(
Background: We evaluated the body fluid (BF) mode on the new Sysmex XE-5000 analyzer.Methods: Red (RBC) and white blood cell (WBC) (differential) counts of BFs (139 patient samples and 87 normal samples) were measured and compared to the Fuchs-Rosenthal chamber and stained cytospin slides.Results: Extended cell counting using the BF mode was noted to have an improved WBC detection limit (CV20%) of 10x10(6)/L. Excellent agreement with the manual method was observed for most BFs [mean bias +2 to 6x10(6)/L for cerebrospinal fluid (CSF) and -1 to 12x10(6)/L for other fluids]. In CSF, the BF-mode counted more WBC (polymorphic nuclear cells) compared with the manual method (mean bias +5 to 6x10(6)/L), especially in samples with low cell counts (