Calibration of the Superconducting Gravimeters SG103, C023, CD029 and CD030.
Calibration of the Superconducting Gravimeters SG103, C023, CD029 and CD030.
复制标题
超导重力计 SG103、C023、CD029 和 CD030 的校准。
DOI:
10.11366/sokuchi1954.47.22
复制
发表时间:
2001
影响因子:
--
通讯作者:
P. Wolf
中科院分区:
文献类型:
--
作者:
R. Falk;M. Harnisch;G. Harnisch;I. Nowak;B. Richter;P. Wolf
In the last years the superconducting gravimeters of the Bundesamt fur Kartographie and Geodasie (BKG) have been calibrated several times by comparison with absolute gravity measurements. Averaging over all measuring campaigns, stable results with errors in the order of 1 nm s 2N are reached. Three of the instruments were also calibrated by the platform method. The internal accuracy of this method is higher by about one order of magnitude. However, there are discrepancies between the results of the two different calibration methods, the origin of which is still not clear.