High-current hollow-cathode source as a radiant intensity standard in the 40-125-nm wavelength range.
High-current hollow-cathode source as a radiant intensity standard in the 40-125-nm wavelength range.
复制标题
高电流空心阴极源作为 40-125 nm 波长范围内的辐射强度标准。
DOI:
10.1364/ao.33.000068
复制
发表时间:
1994
期刊:
影响因子:
1.9
通讯作者:
B. Wende
中科院分区:
文献类型:
--
作者:
J. Hollandt;M. Kuhne;B. Wende
The radiant intensity of VUV emission lines of a high-current hollow-cathode source has been determined for the 40-125-nm spectral range. The source is operated at a constant current of 1 A with an aluminum cathode. Different rare gases are alternatively used as the buffer gas at pressures of ~100 Pa. The radiant intensity has been determined by comparison with the calculable spectral radiant flux of the electron storage ring BESSY. Radiant intensities of the emission lines are in the 7-1400-µW/sr range. The long-term reproducibility of the radiant intensity of the source is within ±10% (2σ value). The systematic uncertainty of the radiometric calibration is better than 9% (√32σ value).