An alternative approach to vector vibrating sample magnetometer detection coil setup
An alternative approach to vector vibrating sample magnetometer detection coil setup
复制标题
矢量振动样品磁力计检测线圈设置的替代方法
DOI:
10.1063/1.1149020
复制
发表时间:
1998
影响因子:
1.6
通讯作者:
J. Lodder
中科院分区:
文献类型:
--
作者:
E. Samwel;T. Bolhuis;J. Lodder
Vector vibrating sample magnetometers (VSMs) can present problems with respect to angular dependent calibration and positional dependency when they are used for measurements on thin film samples, which have dimensions comparable to or larger than the sample–coil distances. The problems are due to the fact that in conventional VSMs the sample is rotating with respect to the coils, when performing angular dependent measurements. In this article a solution is presented based on a setup of VSM detection coils, whose position is linked to that of the sample. Together with a newly designed sample holder, the above mentioned problems are prevented or reduced. The vector detection coil system shows a relatively small error in the determination of the magnetization vector (±1% in the absolute value and ±0.6° in the angle). Furthermore, it has a relatively small positional dependency (1% per mm) combined with a sufficient sensitivity (1 nA m2 or 1 μemu at 10 s time constant) and a capability of using samples up to 10×10 mm2. The improved sample holder for thin film measurements reduces positional problems while, at the same time, reducing the background signals of the holder (to 10 pA m2 per kA/m or 7.958×10−10 emu/Oe).