X-ray CCD calibration system using fluorescent lines
X-ray CCD calibration system using fluorescent lines
复制标题
使用荧光线的 X 射线 CCD 校准系统
DOI:
10.1016/s0168-9002(00)00306-5
复制
发表时间:
2000
影响因子:
1.4
通讯作者:
T. Tsuru
中科院分区:
文献类型:
--
作者:
K. Hamaguchi;Y. Maeda;H. Matsumoto;M. Nishiuchi;H. Tomida;K. Koyama;H. Awaki;T. Tsuru
We have developed a CCD calibration system using fluorescent X-ray lines with energies ranging from 1.49 keV (Al Kα) to 11.2 keV (Se Kα). The absolute X-ray flux is calibrated by a gas proportional counter, while the emerging spectra are monitored by solid-state silicon detectors. In order to suppress contaminating X-rays in the fluorescence spectra, mechanical collimators were set in the X-ray beam line, high-purity targets for fluorescent lines were used, and band-pass filters were put on the X-ray beam line. As for the purity of the fluorescent X-rays, the typical purity achieved was ∼98%.