Thermoelectrically cooled quantum-cascade-laser-based sensor for the continuous monitoring of ambient atmospheric carbon monoxide
Thermoelectrically cooled quantum-cascade-laser-based sensor for the continuous monitoring of ambient atmospheric carbon monoxide
复制标题
DOI:
10.1364/ao.41.001169
复制
发表时间:
2002-02-20
期刊:
影响因子:
1.9
通讯作者:
Allen, MG
中科院分区:
文献类型:
--
作者:
Kosterev, AA;Tittel, FK;Allen, MG
We report the first application of a thermoelectrically cooled, distributed-feedback quantum-cascade laser for continuous spectroscopic monitoring of CO in ambient air at a wavelength of 4.6 mum. A noise-equivalent detection limit of 12 parts per billion was demonstrated experimentally with a 102-cm optical pathlength and a 2.5-min data acquisition time at a 10-kHz pulsed-laser repetition rate. This sensitivity corresponds to a standard error in fractional absorbance of 3 x 10(-5). (C) 2002 Optical Society of America.