TIME CONSTANT IN P300 RECORDING
TIME CONSTANT IN P300 RECORDING
复制标题
DOI:
10.1111/j.1469-8986.1979.tb01440.x
复制
发表时间:
1979-01-01
期刊:
影响因子:
3.7
通讯作者:
DONCHIN, E
中科院分区:
文献类型:
--
作者:
DUNCANJOHNSON, CC;DONCHIN, E
A standard counting task was used to illustrate the effect of the time constant of the recording system on the amplitude and waveform of the P300 component. A random sequence of 1500 Hz (p = .10) and 1000 Hz (p = .90) tones was presented to 1 human male subject. The EEG from CZ and PZ was recorded simultaneously at time constants of 0.1, 0.3, 1.0 and 10.0 s. The ERP [event-related potential] waveforms was markedly affected by the time constant: the shorter the time constant, the smaller the amplitude of both the P300 and slow wave components.