Experimental study of THGEM detector with mini-rim

Experimental study of THGEM detector with mini-rim
复制标题

DOI:
10.1088/1674-1137/36/2/007
复制
发表时间:
2011-05
期刊:
arXiv: Instrumentation and Detectors
影响因子:
--
通讯作者:
Ai-Wu Zhang;Boxiang Yu;Yu-guang Xie;Hongbang Liu;Z. An;Zhi-gang Wang;Xiao Cai;Xilei Sun;F. Shi;Jian Fang;Z. Xue;Qiwen Lu;Li-jun Sun;Yongshuai Ge;Yingbiao Liu;T. Hu;Li Zhou;Junguang Lu
Ai-Wu Zhang;Boxiang Yu;Yu-guang Xie;Hongbang Liu;Z. An;Zhi-gang Wang;Xiao Cai;Xilei Sun;F. Shi;Jian Fang;Z. Xue;Qiwen Lu;Li-jun Sun;Yongshuai Ge;Yingbiao Liu;T. Hu;Li Zhou;Junguang Lu
中科院分区:
其他
文献类型:
--
作者:
Ai-Wu Zhang;Boxiang Yu;Yu-guang Xie;Hongbang Liu;Z. An;Zhi-gang Wang;Xiao Cai;Xilei Sun;F. Shi;Jian Fang;Z. Xue;Qiwen Lu;Li-jun Sun;Yongshuai Ge;Yingbiao Liu;T. Hu;Li Zhou;Junguang Lu

文献摘要

被引文献

相似文献

本文研究了具有小边缘(边缘小于10 μ m)的单、双THGEM探测器(有效面积为5 × 5cm ~ 2)的气体增益和能量分辨率。单、双THGEM探测器的最大增益分别可达5 × 103和2 × 105,而不同孔径和厚度的单、双THGEM探测器对5.9keV X射线的能量分辨率在18%~ 28%之间变化在大气压下,气体(氩气、氖)和少量其它气体(异丁烷、甲烷)。
The gas gain and energy resolution of single and double THGEM detectors (5{\times}5cm2 effective area) with mini-rims (rim is less than 10{\mu}m) were studied. The maximum gain can reach 5{\times}103 and 2{\times}105 for single and double THGEM respectively, while the energy resolution of 5.9 keV X-ray varied from 18% to 28% for both single and double THGEM detectors of different hole sizes and thicknesses.All the experiments were investigated in mixture of noble gases(argon,neon) and small content of other gases(iso-butane,methane) at atmospheric pressure.