Determination of System Geometrical Parameters and Consistency between Scans for Contrast-Enhanced Digital Breast Tomosynthesis
Determination of System Geometrical Parameters and Consistency between Scans for Contrast-Enhanced Digital Breast Tomosynthesis
复制标题
对比增强数字乳腺断层合成系统几何参数和扫描一致性的确定
DOI:
10.1007/978-3-642-31271-7_4
复制
发表时间:
2012
影响因子:
0.6
通讯作者:
Wei Zhao
中科院分区:
文献类型:
--
作者:
D. Scaduto;Wei Zhao
Digital breast tomosynthesis (DBT) requires precise knowledge of acquisition geometry for accurate image reconstruction. Further, image subtraction techniques employed in dual-energy contrast-enhanced tomosynthesis require that scans be performed under nearly identical geometrical conditions. A geometrical calibration algorithm is developed to investigate system geometry and geometrical consistency of image acquisition between consecutive digital breast tomosynthesis scans, according to requirements for dual-energy contrast-enhanced tomosynthesis. Investigation of geometrical accuracy and consistency on a prototype DBT unit reveals accurate angular measurement, but potentially clinically significant differences in acquisition angles between scans. Further, a slight gantry wobble is observed, suggesting the need for incorporation of gantry wobble into image reconstruction, or improvements to system hardware.