Direct sampling of ion.exchanger suspensions for atomic absorption spectrometry with electrothermal atomization
Direct sampling of ion.exchanger suspensions for atomic absorption spectrometry with electrothermal atomization
复制标题
电热雾化原子吸收光谱法中离子交换剂悬浮液的直接采样
DOI:
10.1016/s0003-2670(01)93122-8
复制
发表时间:
1979
期刊:
影响因子:
--
通讯作者:
Z. Slovák
中科院分区:
文献类型:
--
作者:
Z. Slovák
Suspensions of chelating ion-exchange resins with sorbed elements can be injected directly into a Massmann furnace; the reproducibility is similar to that achieved in sampling of solutions. This procedure offers the possibility of preparing standards and samples in the same matrix and so avoiding the usual standardization problems. The technique is discussed in detail for the determination of 1–50 ng of mercury bound on a resin with thiol groups. Methods for the determination of cobalt, nickel and molybdenum (1–16 ng) after sorption on glycolmethacrylate gels with bound 8-hydroxyquinoline are also described.