AMPEROMETRIC L-LACTATE-SENSING ELECTRODE BASED ON A POLYION COMPLEX LAYER CONTAINING LACTATE OXIDASE - APPLICATION TO SERUM AND MILK SAMPLES
AMPEROMETRIC L-LACTATE-SENSING ELECTRODE BASED ON A POLYION COMPLEX LAYER CONTAINING LACTATE OXIDASE - APPLICATION TO SERUM AND MILK SAMPLES
复制标题
DOI:
10.1016/0003-2670(95)00278-8
复制
发表时间:
1995-10-20
影响因子:
6.2
通讯作者:
HIRATA, Y
中科院分区:
文献类型:
--
作者:
MIZUTANI, F;YABUKI, S;HIRATA, Y
An amperometric enzyme electrode for L-lactic acid was prepared by immobilizing lactate oxidase (LOD) in a polyion complex membrane. An aqueous solution containing poly-L-lysine and LOD was placed on a glassy carbon electrode, and an aqueous solution of poly(4-styrenesulfonate) was added to the polycation/LOD mixture and dried. The anodic current (at 1 V vs. Ag/AgCl) of this enzyme electrode increased immediately after the addition of L-lactic acid and the response time was < 5 s. A linear response to L-lactic acid was observed up to 0.3 mM with a detection limit of 0.1 mu M. The enzyme electrode was applied to the determination of L-lactic acid in sera and sour milk: the polyion complex membrane showed permselectivity based on the solute size with the molecular weight cut-off of ca. 100, which was very effective in suppressing the electrochemical interference by L-ascorbic acid, uric acid and acetaminophen. The electrode could be used for at least two months.