How accurate are TheraMon® microsensors at measuring intraoral wear-time? Recorded vs. actual wear times in five volunteers
How accurate are TheraMon® microsensors at measuring intraoral wear-time? Recorded vs. actual wear times in five volunteers
复制标题
DOI:
10.1080/14653125.2017.1365220
复制
发表时间:
2017-01-01
影响因子:
1.3
通讯作者:
Sandler, Jonathan
中科院分区:
文献类型:
--
作者:
Brierley, Catherine A.;Benson, Philip E.;Sandler, Jonathan
Background: The TheraMon (R) microsensor is the most recent device developed to measure the wear-time of removable appliances. The accuracy has not been validated intraorally.Objectives: To determine 1) if the TheraMon (R) microsensor accurately records time when fixed intraorally, and 2) the effect of the intraoral location on the recorded time.Methods: A prospective pilot study, using a convenience sample, was carried out in a UK hospital orthodontic department. Five non-patient volunteers wore microsensors positioned palatal to an upper molar, and buccal to a lower molar for 7 days. Differences between actual amount of wear and the wear recorded by each device were calculated. Differences between sites were also examined.Results: The mean daily wear-time recorded by the upper and lower microsensors combined was 23 hours (95% CI 22.6-23.4), which is a mean under-recording of 4% (CI 2.5-5.8%). The maximum daily under-reporting of wear times was 5.5 hours. Microsensors in the lower buccal sulcus recorded wear-times that were closer to actual wear-times.Conclusions: Assumptions made by the TheraMon (R) microsensors software lead to under-reporting of intraoral wear-time, particularly when placed palatally. These discrepancies could be significant in both clinical practice and research. Adjustment of the microsensor software parameters would improve accuracy, irrespective of the intraoral location.