Fast EPR imaging at 300 MHz using spinning magnetic field gradients
Fast EPR imaging at 300 MHz using spinning magnetic field gradients
复制标题
DOI:
10.1016/j.jmr.2004.02.012
复制
发表时间:
2004-06-01
影响因子:
2.2
通讯作者:
Zweier, JL
中科院分区:
文献类型:
--
作者:
Deng, YM;He, GL;Zweier, JL
Electron paramagnetic resonance imaging (EPRI) technology has rapidly progressed in the last decade enabling many important applications in the fields of biology and medicine. At frequencies of 300-1200 MHz a range of in vivo applications have been performed. However, the requisite imaging time duration to acquire a given number of projections, limits the use of this technique in many in vivo applications where relatively rapid kinetics occur. Therefore, there has been a great need to develop approaches to accelerate EPRI data acquisition. We report the development of a fast low-frequency EPRI technique using spinning magnetic field gradients (SMFG). Utilizing a 300 MHz CW (continuous wave) EPRI system, SMFG enabled over 10-fold accelerated acquisition of image projections. 2D images with over 200 projections could be acquired in less than 3 s and with 20 s acquisitions good image quality was obtained on large aqueous free radical samples. This technique should be particularly useful for in vivo studies of free radicals and their metabolism. (C) 2004 Elsevier Inc. All rights reserved.