New Technique for Determining Instantaneous Myocardial Force‐Velocity Relations in the Intact Heart

New Technique for Determining Instantaneous Myocardial Force‐Velocity Relations in the Intact Heart
复制标题

确定完整心脏瞬时心肌力-速度关系的新技术

DOI:
--
复制
发表时间:
1971
影响因子:
20.1
通讯作者:
L. Dexter
L. Dexter
中科院分区:
医学1区
文献类型:
--
作者:
W. Grossman;H. Brooks;S. Meister;H. Sherman;L. Dexter

文献摘要

被引文献

相似文献

一种新的技术,描述了心肌力-速度关系的瞬时测定。该方法采用心室内压对数的电子微分,产生(dP/dt)P−1(心室压上升速率[dP/dt]与同时心室压[P]的比值)的连续在线记录。另一种技术描述了力-速度矢量环的自动投影,以逐搏方式在纵坐标上显示(dP/dt)P−1,在横坐标上显示心室压力。通过常规方法测定的(dP/dt)P−1与通过使用对数放大器电路以电子方式导出的(dP/dt)P−1之间具有良好的相关性(r=0.982)。描述了实验研究,其记录了通过本方法确定的(dP/dt)P−1对已知影响心肌收缩性的干预的响应性。在输注CaCl 2、去甲肾上腺素和胰高血糖素后观察到(dP/dt)P−1增加,而在戊巴比妥后观察到下降。Inclusions(dP/dt)P−1是心肌收缩力的有效指标,本方法允许在各种情况下对心室功能的变化进行在线逐搏评估。
A new technique is described for the instantaneous determination of myocardial force-velocity relationships. The method employs electronic differentiation of the logarithm of intraventricular pressure, which yields a continuous on-line record of (dP/dt)P−1 (the ratio of the rate of rise of ventricular pressure [dP/dt] to the simultaneous ventricular pressure [P]). A further technique is described for the automatic projection of force-velocity vector loops, displaying (dP/dt)P−1 on the ordinate against ventricular pressure on the abscissa in a beat-to-beat fashion. An excellent correlation (r=0.982) was demonstrated between (dP/dt)P−1 determined by conventional methods and that derived electronically by use of the logarithmic amplifier circuit. Experimental studies are described which document the responsiveness of (dP/dt)P−1 determined by the present method to interventions known to affect myocardial contractility. An increase in (dP/dt)P−1 was observed following infusions of CaCl2, norepinephrine, and glucagon, and a decrease following pentobarbital. Insofar as (dP/dt)P−1 is a valid index of myocardial contractility, the present method permits on-line, beat-to-beat evaluation of changes in ventricular function under a variety of circumstances.