Laser-patterned lateral-flow devices with multiple flow-paths for semi-quantitative measurement of the inflammatory biomarker, C-reactive protein (Conference Presentation)
Laser-patterned lateral-flow devices with multiple flow-paths for semi-quantitative measurement of the inflammatory biomarker, C-reactive protein (Conference Presentation)
复制标题
具有多个流路的激光图案侧流装置,用于半定量测量炎症生物标志物 C 反应蛋白(会议演示)
DOI:
10.1117/12.2542692
复制
发表时间:
2020
期刊:
影响因子:
--
通讯作者:
Iles A
中科院分区:
文献类型:
--
作者:
Iles A
We report a novel solution for semi-quantitative detection of C-reactive protein (CRP), a common inflammatory marker, using a multi-path lateral-flow device (LFD) fabricated via the precise partitioning of the single flow-path of a standard LFD using a laser direct-write technique. The multiple isolated flow-paths allow detection of different concentrations by individually setting different cut-off levels. The semi-quantitative readout is observed via simple appearance/absence of test lines without the need for a reader or further analyses. We have successfully demonstrated the use of these multi-path LFDs for semi-quantitative detection of CRP and will present validation in human samples.