Verifying Position Errors In CNC Gear Measuring InstrumentsUsing A Laser Interferometer With Dynamic Data Capture Software

Verifying Position Errors In CNC Gear Measuring InstrumentsUsing A Laser Interferometer With Dynamic Data Capture Software
复制标题

使用带有动态数据捕获软件的激光干涉仪验证 CNC 齿轮测量仪器中的位置误差

DOI:
10.2495/lamdamap010311
复制
发表时间:
1970
期刊:
WIT transactions on engineering sciences
影响因子:
--
通讯作者:
D. Wehmeyer
D. Wehmeyer
中科院分区:
--
文献类型:
--
作者:
R. Bicker;R. Frazer;D. Wehmeyer

文献摘要

被引文献

相似文献

CNC gear measuring instruments measure gears by scanning across the tooth surface. A Renishaw laser interferometer system with position based dynamic data capture hardware and software was used to verify position errors on slides at the normal scanning speed used when measuring gears. The results from a Hofler EMZ 632 CNC gear tester in the UK national Gear Metrology Laboratory are presented and show that the techmque generates realistic data for calibration purposes under closely controlled laboratory conditions. A comparison between data gathered under dynamic conditions and static conditions is made. Some important considerations applying the technique are discussed.