Knitted ECG Electrodes in Relaxed Fitting Garments
Knitted ECG Electrodes in Relaxed Fitting Garments
复制标题
宽松服装中的针织心电图电极
DOI:
10.1109/jsen.2023.3236723
复制
发表时间:
2023
影响因子:
4.3
通讯作者:
K. Thielemans
中科院分区:
文献类型:
--
作者:
K. Fobelets;Ghena Hammour;K. Thielemans
A wide range of signal quality indices (SQIs) related to statistical methods are used to guide the optimization process of knitted textile electrodes for electrocardiogram (ECG) recordings. The electrode structure and composition as well as their integration into a garment are evaluated in view of a fully knitted garment. The dry electrodes with the best SQIs are obtained by using conductive yarn only with a compact knit structure and a medium level of roughness. The best SQIs for the e-garment were obtained by sewing electrodes at their edges only, into the knitted garment. This implementation outperforms the intarsia and double-knit method as it allows the garment some independent movement from the electrodes, reducing motion artifacts. Tests done on a healthy volunteer demonstrate excellent system performance under gentle ambulation. The advantage of using SQIs in the optimization process of dry textile ECG electrodes is that they offer a quantitative benchmark against which to compare other approaches. The fully knitted clothing allows for more relaxed e-garments when gentle ambulation is considered.