External fixator clamp reuse degrades clamp mechanical performance.

External fixator clamp reuse degrades clamp mechanical performance.
复制标题

外部固定器夹具的重复使用会降低夹具的机械性能。

DOI:
10.1111/j.1532-950x.2009.00521.x
复制
发表时间:
2009
期刊:
Veterinary surgery : VS
影响因子:
--
通讯作者:
Soslowsky,LouisJ
Soslowsky,LouisJ
中科院分区:
--
文献类型:
--
作者:
Gilley,RobertS;Beason,DavidP;Snyder,DavidM;Boston,RaymondC;Radin,Alex;Kapatkin,AmyS;Smith,GailK;Soslowsky,LouisJ

文献摘要

参考文献

被引文献

相似文献

Objective—To determine the effects of clamp reuse for the Kirschner–Ehmer (KE); Securos; and the IMEX–SK clamp.Study Design—Experimental bench test of mechanical properties.Methods—Specially designed fixtures were used to mechanically test 18 clamps of each type with respect to 6 mechanical variables: fixator pin slippage, connecting bar slippage, fixator pin rotation, connecting bar rotation, and clockwise and counterclockwise clamp–bolt axis pivot. Each clamp was tested 6 times for each variable at 7.68 Nm of clamp–bolt tightening torque. Results were compared using repeated measures ANOVA.Results—For the IMEX–SK clamp, a significant degradation in the force required to cause slippage was found for connecting bar slippage and fixator pin rotation; however, this clamp also had a significant increase in the force to initiate slippage for clockwise clamp–bolt axis pivot with reuse. The Securos clamp had significant degradation in connecting bar slippage, connecting bar rotation, and fixator pin rotation whereas the KE clamp had significant degradation in connecting bar slippage only.Conclusions—All 3 external fixator clamp types degraded in 1 or more movement variables in their ability to resist motion with reuse. The IMEX and Securos clamps were more subject to degradation than the KE clamp and this may have clinical importance for fixator composite rigidity.Clinical Relevance—Fracture‐reduction stability is related to the ability to resist motion within a clamp. The unpredictable nature of degradation we found cautions against repeated use. Clinicians should consider reuse of external fixator clamps with the knowledge that repeated use degrades clamp mechanical performance.
螺栓拧紧扭矩、钢丝尺寸和部件重复使用对圆形外固定中钢丝固定的影响。
DOI: --
发表时间: 2002
期刊: Veterinary surgery
影响因子: 1.8
作者:
M. A. Wosar;D. Marcellin;S. Roe
通讯作者: S. Roe
DOI: --
发表时间: 1993
期刊: Veterinary surgery
影响因子: 1.8
作者:
W. Dernell;J. Harari;D. Blackketter
通讯作者: D. Blackketter
评估在犬尸体股骨中钻孔和不钻孔的情况下放置的无螺纹和增强型螺纹销的皮质骨损伤和轴向保持力。
DOI: --
发表时间: 1996
期刊: Journal of the American Veterinary Medical Association
影响因子: --
作者:
Anderson Ma;Fred A. Mann;Kinden Da;Colette C. Wagner
通讯作者: Colette C. Wagner
DOI: --
发表时间: 1992
期刊: The Veterinary clinics of North America. Small animal practice
影响因子: --
作者:
Ross H. Palmer;DONALD A. Hulse;W. Hyman;Darcie R. Palmer
通讯作者: Darcie R. Palmer
狗 I 型外部骨骼固定器中使用的非螺纹、增强型螺纹和 Ellis 固定销的比较。
DOI: --
发表时间: 1993
期刊: Veterinary surgery
影响因子: 1.8
作者:
M. Anderson;F. Mann;C. Wagner;A. W. Hahn;B. Jiang;J. Tomlinson
通讯作者: J. Tomlinson