Attempt at visualizing breast cancer with X-ray dark field imaging
Attempt at visualizing breast cancer with X-ray dark field imaging
复制标题
DOI:
10.1143/jjap.44.l528
复制
发表时间:
2005-01-01
期刊:
影响因子:
--
通讯作者:
Hirano, M
中科院分区:
文献类型:
--
作者:
Ando, M;Yamasaki, K;Hirano, M
X-ray dark-field imaging (DFI) can clearly visualize breast cancer phantoms and cancer cell nests, stroma, fat tissue, ductus lactiferi, muscle, collagen fibers at stroma and calcification in a 2.8-mm-thick breast cancer pathological specimen. The system comprises a Bragg asymmetric-cut monochro-collimator and a 2.124-mm-thick Si 440 Laue diffraction analyzer at 35 keV. Both optical elements are Floating Zone made silicon crystals. The view size of 33 mm (H) x 19.5 mm (V) and the spatial resolution of 10 mu m or better are obtainable at the vertical wiggler beamline BL14B at the Photon Factory.