NITROGEN ANALYZER ADAPTATION FOR USE IN PRESENCE OF HALOGENATED ANESTHETICS
NITROGEN ANALYZER ADAPTATION FOR USE IN PRESENCE OF HALOGENATED ANESTHETICS
复制标题
DOI:
10.1097/00000542-197512000-00014
复制
发表时间:
1975-01-01
期刊:
影响因子:
8.8
通讯作者:
CRAIG, DB
中科院分区:
文献类型:
--
作者:
GILMOUR, IJ;HOLLAND, R;CRAIG, DB
Interposition of activated carbon between a nitrogen analyzer needle valve and gas mixtures containing halothane, methoxyflurane, enflurane, or isoflurane eliminated the interfering effects of these agents, but added sampling delays. Placement of a carbon-filled Swinney adaptor in the sampling line between the needle valve and analyzer also eliminated the interference. Linearity of response (range 0 to 80 per cent N2) and sampling delay (40 msec) were unchanged. The 90 per cent fall time (step change 80 to 0 per cent N2) increased from 35 to 100 msec. This adaptation allows rapid analysis of nitrogen concentrations during general anesthesia.