Influence of matrix solution conditions on the MALDI-MS analysis of peptides and proteins
Influence of matrix solution conditions on the MALDI-MS analysis of peptides and proteins
复制标题
DOI:
10.1021/ac9507956
复制
发表时间:
1996-01-01
影响因子:
7.4
通讯作者:
Chait, BT
中科院分区:
文献类型:
--
作者:
Cohen, SL;Chait, BT
Sample-matrix preparation procedures are shown to greatly influence the quality of the matrix-assisted laser desorption/ionization (MALDI) mass spectra of peptides and proteins, In particular, dramatic mass discrimination effects are observed when the matrix 4-hydroxy-alpha-cyanocinnamic acid is used for analyzing complex mixtures of peptides and proteins. The discrimination effects are found to be strongly dependent on the sample-matrix solution composition, pH, and the rates at which the sample-matrix cocrystals are grown. These findings demonstrate the need to exercise great care in performing and interpreting the MALDI analysis of biological samples, The results also indicate that there is a reverse-phase chromatographic-like dimension in the sample-matrix preparation procedures that can be exploited to optimize the analysis. The present work describes the conditions under which the majority of components of a complex mixture of peptides and proteins can be successfully measured.