The role of capsule on the performance of a dry powder inhaler using computational and experimental analyses
The role of capsule on the performance of a dry powder inhaler using computational and experimental analyses
复制标题
DOI:
10.1007/s11095-005-4587-y
复制
发表时间:
2005-06-01
影响因子:
3.7
通讯作者:
Raper, JA
中科院分区:
文献类型:
--
作者:
Coates, MS;Fletcher, DF;Raper, JA
Purpose. To study the fundamental effects of the spinning capsule on the overall performance of a dry powder inhaler (Aerolizer(R)).Methods. The capsule motion was visualized using high-speed photography. Computational fluid dynamics (CFD) analysis was performed to determine the flowfield generated in the device with and without the presence of different sized capsules at 60 l min(-1). The inhaler dispersion performance was measured with mannitol powder using a multistage liquid impinger at the same flowrate.Results. The capsule size (3, 4, and 5) was found to make no significant difference to the device flowfield, the particle-device impaction frequency, or the dispersion performance of the inhaler. Reducing the capsule size reduced only the capsule retention by 4%. In contrast, without the presence of the spinning capsule, turbulence levels were increased by 65%, FPFEm (wt% particles