Evaluation of radio frequency microcoils as nuclear magnetic resonance detectors in low-homogeneity high-field superconducting magnets
Evaluation of radio frequency microcoils as nuclear magnetic resonance detectors in low-homogeneity high-field superconducting magnets
复制标题
DOI:
10.1063/1.1149201
复制
发表时间:
1998-11
影响因子:
1.6
通讯作者:
A. Wright;T. A. Neideen;R. Magin;James A. Norcross
中科院分区:
文献类型:
--
作者:
A. Wright;T. A. Neideen;R. Magin;James A. Norcross
We describe here experiments evaluating the performance of solenoidal radio frequency probes having submillimeter dimensions (microcoils) as detectors for liquid nuclear magnetic resonance (NMR) in very low-homogeneity (100 ppm/cm) magnetic fields. Performance is based on the measured H2O linewidth. A series of solenoidal microcoils having sample volumes 8, 53, and 593 nl were filled with distilled H2O and evaluated for smallest obtainable unshimmed NMR spectral linewidths in a vertical bore superconducting magnet, stabilized at 5.9 T (1H frequency=250 MHz). The smallest microcoil (472 μm diameter) gave a smallest H2O linewidth of 525 Hz, 25 times smaller than that from a standard 5.7 mm probe. Linewidth increased approximately as the square root of sample volume. For comparison, shimmed H2O linewidths using the same microcoils in a high-homogeneity (0.1 ppm/cm) NMR magnet were also measured. Shimmed linewidths in the high-homogeneity magnet were two orders of magnitude smaller and exhibited a similar dep...