First intercomparison of Tn-Rn discriminative measurements in China.
First intercomparison of Tn-Rn discriminative measurements in China.
复制标题
DOI:
10.1093/rpd/ncm233
复制
发表时间:
2007-06
影响因子:
1
通讯作者:
S. Qiu;D. Xiao;L. Liu
中科院分区:
文献类型:
--
作者:
S. Qiu;D. Xiao;L. Liu
The intercomparison measurements with Tn-Rn discriminative monitor were conducted at the Radon Laboratory of Nanhua University. The intercomparison results indicates that the Tn-Rn discrimination of most monitors (including the solid-state nuclear track detector of participant A and participant B, continuous monitors of RAD7 and RTM2100) keeps well in a wide range of the ratio of Tn to Rn. There are great differences of Tn measurement results among RAD7, RTM2100 and reference instrument. The discussion on the reason of these differences reveals the necessity of improvement on monitor calibration, Tn source make and the necessity of more international intercomparisons.