Rate-independent characteristics of an arthroscopically implantable force probe in the human Achilles tendon
Rate-independent characteristics of an arthroscopically implantable force probe in the human Achilles tendon
复制标题
DOI:
10.1016/s0021-9290(98)00167-5
复制
发表时间:
1999-02-01
影响因子:
2.4
通讯作者:
Hale, JE
中科院分区:
文献类型:
--
作者:
Hall, GW;Crandall, JR;Hale, JE
The effect of loading rate on specimen calibration was investigated for an implantable force sensor of the two-point loading variety. This variety of sensor incorporates a strain gage to measure the compressive load applied to the sensor due Co tensile loading in a soft tissue specimen. The Achilles tendon in each of four human cadaveric lower extremities was instrumented with a force sensor and then loaded in tension using a materials testing machine. Each specimen was tensile tested at three different displacement rates, 0.25, 2.5 and 12.7 cm s(-1), corresponding with mean loading rates of 33.8, 513.2, and 2838.6 N s(-1), respectively. A calibration curve relating the force sensor signal and applied tendon tension was generated for each specimen/displacement rate combination. For each specimen, calibration curves were compared by calculating an RMS error for the entire data set (e(RMS) = 1.6% of the full load value) and a coefficient of determination, R-2, of a curve fit through all of the data (R-2 = 99.6%). Over the range of rates tested, no measurable change in sensor sensitivity due to loading rate was observed. Hysteresis for all displacement rates was on the order of 2.4%. (C) 1999 Elsevier Science Ltd. All rights reserved.