Manipulator-driven selection of semi-active MR-visible markers
Manipulator-driven selection of semi-active MR-visible markers
复制标题
机械手驱动的半主动 MR 可见标记选择
DOI:
10.1002/rcs.1846
复制
发表时间:
2018
期刊:
影响因子:
--
通讯作者:
Tsekos, Nikolaos V.
中科院分区:
文献类型:
--
作者:
An, Junmo;Webb, Andrew G.;Shah, Dipan J.;Chin, Karen;Tsekos, Nikolaos V.
BackgroundA method for the identification of semi‐active fiducial magnetic resonance (MR) markers is presented based on selectively optically tuning and detuning them.MethodsFour inductively coupled solenoid coils with photoresistors were connected to light sources. A microcontroller timed the optical tuning/detuning of coils and image collection. The markers were tested on an MR manipulator linking the microcontroller to the manipulator control to visibly select the marker subset according to the actuated joint.ResultsIn closed‐loop control, the average and maximum were 0.76° ± 0.41° and 1.18° errors for a rotational joint, and 0.87 mm ± 0.26 mm and 1.13 mm for the prismatic joint.ConclusionsThis technique is suitable for MR‐compatible actuated devices that use semi‐active MR‐compatible markers.