Four-channel optically pumped atomic magnetometer for magnetoencephalography
Four-channel optically pumped atomic magnetometer for magnetoencephalography
复制标题
DOI:
10.1364/oe.24.015403
复制
发表时间:
2016-07-11
期刊:
影响因子:
3.8
通讯作者:
Schwindt, Peter D. D.
中科院分区:
文献类型:
--
作者:
Colombo, Anthony P.;Carter, Tony R.;Schwindt, Peter D. D.
We have developed a four-channel optically pumped atomic magnetometer for magnetoencephalography (MEG) that incorporates a passive diffractive optical element (DOE). The DOE allows us to achieve a long, 18-mm gradiometer baseline in a compact footprint on the head. Using gradiometry, the sensitivities of the channels are < 5 fT/Hz(1/2), and the 3-dB bandwidths are approximately 90 Hz, which are both sufficient to perform MEG. Additionally, the channels are highly uniform, which offers the possibility of employing standard MEG post-processing techniques. This module will serve as a building block of an array for magnetic source localization. (C) 2016 Optical Society of America