First images of a digital autoradiography system based on a Medipix2 hybrid silicon pixel detector
First images of a digital autoradiography system based on a Medipix2 hybrid silicon pixel detector
复制标题
DOI:
10.1088/0031-9155/48/12/403
复制
发表时间:
2003-06-21
影响因子:
3.5
通讯作者:
Russo, P
中科院分区:
文献类型:
--
作者:
Mettivier, G;Montesi, MC;Russo, P
We present the first images of beta autoradiography obtained with the high-resolution hybrid pixel detector consisting of the Medipix2 single photon counting read-out chip bump-bonded to a 300 mum thick silicon pixel detector. This room temperature system has 256 x 256 square pixels of 55 mum pitch (total sensitive area of 14 x 14 mm 2), with a double threshold discriminator and a 13-bit counter in each pixel. It is read out via a dedicated electronic interface and control software, also developed in the framework of the European Medipix2 Collaboration. Digital beta autoradiograms of C-14 microscale standard strips (containing separate bands of increasing specific activity in the range 0.0038-32.9 kBq g(-1)) indicate system linearity down to a total background noise of 1.8 x 10(-3) counts mm(-2) s(-1). The minimum detectable activity is estimated to be 0.012 Bq for 36 000 s exposure and 0.023 Bq for 10 800 s exposure. The measured minimum detection threshold is less than 1600 electrons (equivalent to about 6 keV Si). This real-time system for beta autoradiography offers lower pixel pitch and higher sensitive area than the previous Medipix 1-based system. It has a C-14 sensitivity better than that of micro channel plate based systems, which, however, shows higher spatial resolution and sensitive area.