Experimental evaluation of pencil beam algorithm by measurements of dose distributions of protons traversing an L-shaped phantom
Experimental evaluation of pencil beam algorithm by measurements of dose distributions of protons traversing an L-shaped phantom
复制标题
DOI:
10.1143/jjap.40.441
复制
发表时间:
2001-01-01
期刊:
影响因子:
--
通讯作者:
Yasuoka, K
中科院分区:
文献类型:
--
作者:
Kohno, R;Takada, Y;Yasuoka, K
A proton dose-calculation method has been developed on the basis of the pencil-beam algorithm (PBA), which is expected to improve the precision of the conventional method based on the broad-beam algorithm (BBA). In order to verify the accuracy of calculation by the PEA, dose distributions formed by a horizontal proton beam traversing a phantom with an L-shaped horizontal cross section were measured using a silicon semiconductor detector (SSD) in a water vessel. The results of the measured dose distributions agree well with the ones calculated casing the PEA within an rms error of 2.3%. Therefore, the dose-calculation method by the PEA is useful and applicable to actual treatment planning of the proton therapy.