Multi-Echo Acquisition of MR Angiography and Venography of the Brain at 3 Tesla
Multi-Echo Acquisition of MR Angiography and Venography of the Brain at 3 Tesla
复制标题
DOI:
10.1002/jmri.21833
复制
发表时间:
2009-08-01
影响因子:
4.4
通讯作者:
Tanabe, Jody
中科院分区:
文献类型:
--
作者:
Du, Yiping P.;Jin, Zhaoyang;Tanabe, Jody
Purpose: To improve the visibility of veins in susceptibility-weighted imaging (SWI) using a multi-gradient echo acquisition.Materials and Methods: A three-dimensional multi-echo gradient-echo pulse sequence was developed for simultaneous acquisition of MR angiography and multiple volumes of MR venography (MRV) of the brain. The first echo was acquired for MR angiography using the time-of-flight inflow effect. The subsequent echoes were acquired for SWI-based MRV at different echo times (TEs).Results: Multiple MRV datasets acquired at different TEs were complementary in depicting the venous vasculature. MRV data acquired at a longer TE had a higher venous contrast and stronger susceptibility weighting, whereas MRV data acquired at a shorter TE had a higher signal-to-noise ratio and less severe off-resonance artifacts. Three-dimensional mapping of local field gradients was calculated and the transverse relaxivity (R(2)*) at each voxel was quantified using multi-TE exponential fitting.Conclusion: Multi-echo acquisition of MR angiography and venography demonstrated improved visibility of venous vasculature especially in regions with severe field inhomogeneity compared with conventional acquisition of. SWI and dual-echo acquisition of MR angiography and venography.