Pediatric vision screening using the plusoptiX A12C photoscreener in Chinese preschool children aged 3 to 4 years.
Pediatric vision screening using the plusoptiX A12C photoscreener in Chinese preschool children aged 3 to 4 years.
复制标题
使用 plusoptiX A12C 光电筛查仪对中国 3 至 4 岁学龄前儿童进行儿童视力筛查
DOI:
10.1038/s41598-017-02246-6
复制
发表时间:
2017-05-17
影响因子:
4.6
通讯作者:
Liu H
中科院分区:
文献类型:
--
作者:
Huang D;Chen X;Zhang X;Wang Y;Zhu H;Ding H;Bai J;Chen J;Fu Z;Wang Z;Liu H
This study evaluated the performance of plusoptiX A12C in detecting amblyopia risk factors (ARFs) in Chinese children aged 3-to-4-year. PlusoptiX examination was successfully conducted among 1,766 subjects without cycloplegia to detect refractive error, asymmetry and media opacity. Cycloplegic retinoscopy (CR) was conducted on 357 children suspected of having vision abnormalities. Statistical differences between CR and the device were confirmed using the mean spherical value (+1.41 ± 0.87 D versus +1.14 ± 0.81 D), cylindrical value (−0.47 ± 0.64 versus −0.84 ± 0.78) and spherical equivalent (SE) value (+1.17 ± 0.84 D versus +0.72 ± 0.64 D) (allP< 0.0001). In the emmetropia group, the differences were statistically significant for the cylinder and SE (allP< 0.0001) but not the sphere (P= 0.33). In the hyperopia group, the differences were statistically significant for the sphere, cylinder and SE (allP< 0.0001). For refractive and strabismic ARFs detection, the sensitivity, specificity, positive predictive value, and negative predictive value were calculated, respectively.