Comparison of 3 Optical Navigation Systems for Computer-Aided Maxillofacial Surgery
Comparison of 3 Optical Navigation Systems for Computer-Aided Maxillofacial Surgery
复制标题
DOI:
10.1001/archotol.134.10.1080
复制
发表时间:
2008-10-01
影响因子:
--
通讯作者:
Metzger, Marc Christian
中科院分区:
文献类型:
--
作者:
Strong, E. Bradley;Rafii, Amir;Metzger, Marc Christian
Objective: To compare the accuracy of 3 computeraided surgery systems for maxillofacial reconstruction.Design: Evaluation of 3 computer-aided surgery systems: StealthStation, VectorVision, and Voxim.Setting: The University of California, Davis, Department of Otolaryngology computer-aided surgery laboratory.Participants: Four fresh cadaveric heads.Main Outcome Measure: Mean target registration error.Results: The StealthStation was the most accurate (mean [SD] target registration error, 1.00 [0.04] mm), followed by VectorVision (1.13 [0.05] mm) and then Voxim (1.34 [0.04] mm). All values met statistical significance (P < .05).Conclusions: Measurable accuracy differences were found among the navigation systems evaluated. The StealthStation was the most accurate. However, the differences are small, and the clinical significance for maxillofacial reconstruction is negligible.