Assessment of technical and biological parameters of volumetric quantitative computed tomography of the foot: a phantom study.
Assessment of technical and biological parameters of volumetric quantitative computed tomography of the foot: a phantom study.
复制标题
DOI:
10.1007/s00198-011-1851-3
复制
发表时间:
2012-07
影响因子:
4
通讯作者:
Prior, F. W.
中科院分区:
文献类型:
--
作者:
Smith, K. E.;Whiting, B. R.;Reiker, G. G.;Commean, P. K.;Sinacore, D. R.;Prior, F. W.
关键词:
Few studies exist for bone densitometry of the whole foot. A phantom study demonstrated the sources of error and necessary controls for accurate quantitative computed tomography of the foot. A loss in bone mineral density in the small foot bones may be an early indicator of diabetic foot complications. Volumetric quantitative computed tomography (vQCT) facilitates assessment of pedal bone osteopenia, which in the presence of peripheral neuropathy may well be an early sign of diabetic foot deformity. To date, sources and magnitudes of error in foot vQCT measurements have not been reported. Foot phantoms were scanned using a 64-slice CT scanner. Energy (kVp), table height, phantom size and orientation, location of “bone” inserts, insert material, location of calibration phantom, and reconstruction kernel were systematically varied during scan acquisition. Energy (kVp) and distance from the isocenter (table height) resulted in relative attenuation changes from −5% to 22% and −5% to 0%, respectively, and average bone mineral density (BMD) changes from −0.9% to 0.0% and −1.1% to 0.3%, respectively, compared to a baseline 120 kVp scan performed at the isocenter. BMD compared to manufacturer specified values ranged on average from −2.2% to 0.9%. Phantom size and location of bone-equivalent material inserts resulted in relative attenuation changes of −1.2% to 1.4% compared to the medium sized phantom. This study demonstrated that variations in kVp and table height can be controlled using a calibration phantom scanned at the same energy and height as a foot phantom; however, error due to soft tissue thickness and location of bones within a foot cannot be controlled using a calibration phantom alone.
登录
查看更多内容
影响因子:
4.8
作者:
Bligh, Michael;Bidaut, Luc;Cody, Dianna D.
通讯作者:
Cody, Dianna D.
影响因子:
2.7
作者:
Schon, LC;Weinfeld, SB;Resch, S
通讯作者:
Resch, S
影响因子:
0.6
作者:
Chantelau, Ernst;Kimmerle, Renate;Poll, Ludger W
通讯作者:
Poll, Ludger W
影响因子:
1.3
作者:
GLUER, CC;REISER, UJ;GENANT, HK
通讯作者:
GENANT, HK
影响因子:
3.8
作者:
Hangartner, Thomas N.;Short, David F.
通讯作者:
Short, David F.