Socket/stump interface dynamic pressure distributions recorded during the prosthetic stance phase of gait of a trans-tibial amputee wearing a hydrocast socket
Socket/stump interface dynamic pressure distributions recorded during the prosthetic stance phase of gait of a trans-tibial amputee wearing a hydrocast socket
复制标题
佩戴 Hydrocast 承窝的经胫骨截肢者步态假肢站立阶段记录的承窝/残肢界面动态压力分布
DOI:
--
复制
发表时间:
1999
影响因子:
1.5
通讯作者:
A. Buis
中科院分区:
文献类型:
--
作者:
P. Convery;A. Buis
Force sensing resistors (FSR) have been used to measure dynamic stump/socket interface pressures during the gait of a trans-tibial amputee. A total of 350 pressure sensors were attached to the inner wall of a hydrocast socket. Data were sampled at 150Hz during approximately 0.8 seconds of prosthetic stance of gait. The dynamic pressure distributions within a hand cast socket reported by Convery and Buis (1998) are compared with those monitored within a hydrocast socket for the same amputee. The pressure gradients within the hydrocast socket are less than that of the hand cast Patellar-Tendon-Bearing (PTB) socket. The proximal “ring” of high pressure in the hand cast PTB socket is replaced with a more distal pressure in the hydrocast socket.