Flow dissolution of 1,5-diphenylcarbazide for the determination of chromium(VI)
Flow dissolution of 1,5-diphenylcarbazide for the determination of chromium(VI)
复制标题
DOI:
10.1039/an9962100613
复制
发表时间:
1996
期刊:
影响因子:
4.2
通讯作者:
F. Andrade;M. Tudino;O. Troccoli
中科院分区:
文献类型:
--
作者:
F. Andrade;M. Tudino;O. Troccoli
An alternative procedure for the flow injection spectrophotometric determination of CrVI with 1,5-diphenylcarbazide (DPC), involving on-line dissolution of the reagent, is presented. Solid DPC (25% m/m in silica) is packed in a PTFE column, which is then incorporated in a single-line flow system so that reagent dissolution is effected as the sample passes through the column. This procedure reduces the chemical contribution to the dispersion value to a minimum and gives a low detection limit owing to the absence of refractive index peaks. An improvement of the figures of merit compared with those reported previously is shown. The calibration graph is linear up to 0.800 mg l–1 CrVI, with a slope of 0.49 l mg–1; the detection limit is 0.005 mg l–1. The relative standard deviation is better than 1%.