Concordance between direct and indirect measurements of free 25-OH vitamin D.

Concordance between direct and indirect measurements of free 25-OH vitamin D.
复制标题

DOI:
10.1016/j.cca.2017.10.016
复制
发表时间:
2017-12
期刊:
Clinica chimica acta; international journal of clinical chemistry
影响因子:
--
通讯作者:
P. Peris;X. Filella;A. Monegal;N. Guañabens;L. Foj;M. Bonet;D. Boquet;E. Casado;D. Cerdá
P. Peris;X. Filella;A. Monegal;N. Guañabens;L. Foj;M. Bonet;D. Boquet;E. Casado;D. Cerdá
中科院分区:
其他
文献类型:
--
作者:
P. Peris;X. Filella;A. Monegal;N. Guañabens;L. Foj;M. Bonet;D. Boquet;E. Casado;D. Cerdá

文献摘要

被引文献

相似文献

目前,比较游离维生素D的定量测定方法(直接测定法[direct 25-OHDF]和计算估计法[calculate 25-OHDF])的数据很少。本研究的目的是分析这两种25- ohdf分析方法(直接与计算)之间的一致性。方法对173例35 ~ 45岁健康女性血清总25-OHDT、维生素D结合蛋白(DBP) (R&D Systems ELISA)、计算25-OHDF(DBP、白蛋白和25-OHDT)和直接25-OHDF(DIAsource ELISA)进行测定。采用Bland-Altman图和总偏差指数(TDI)评价一致性。结果25-OHDFin计算值和直接值的平均值分别为5.27±2.5 pg/mL和3.83±1.01 pg/mL。我们发现,与使用的方法无关,与维生素D缺乏症患者和非维生素D缺乏症患者相比,25-OHDFon的值显著降低。由Bland-Altman图评估的总偏差指数显示两个测量值的一致性较低。只有低25- ohdf水平是一致的。结论25- ohdf两种分析方法的一致性较低,存在浓度依赖偏倚。需要进一步的研究来明确25- ohdf测量的参考值和适应症。
At present, data comparing the quantification methods for measurement of free vitamin D (direct assay [direct 25-OHDF] and estimated by calculation [calculated 25-OHDF]), are scarce. The aim of this study was to analyse the concordance between these two methods of 25-OHDFanalysis (direct vs. calculated).MethodsSerum values of total 25-OHD (25-OHDT), vitamin D binding protein (DBP) (by R&D Systems ELISA), calculated 25-OHDF(by DBP, albumin and 25-OHDT) and direct 25-OHDF(by DIAsource ELISA) were analysed in 173 healthy women (aged 35–45 years). Assessment of concordance was evaluated by the Bland-Altman plot and the total deviation index (TDI).ResultsThe mean values of calculated and direct 25-OHDFin these subjects were 5.27 ± 2.5 and 3.83 ± 1.01 pg/mL, respectively. We found significantly lower values of 25-OHDFon comparing subjects with and without vitamin D deficiency, independently of the method used. The total deviation index evaluated by the Bland-Altman plot showed low concordance for both measurements. Only low 25-OHDFlevels were concordant.ConclusionsThis study shows that the concordance between these two methods of 25-OHDFanalysis is low and has a concentration dependent bias. Further studies are necessary to clarify the reference values and the indications for 25-OHDFmeasurement.