MEASUREMENT OF TIDAL FLOW USING A TRANSIT-TIME ULTRASONIC BREATH ANALYZER
MEASUREMENT OF TIDAL FLOW USING A TRANSIT-TIME ULTRASONIC BREATH ANALYZER
复制标题
DOI:
10.1111/j.1365-2044.1995.tb05999.x
复制
发表时间:
1995-05-01
期刊:
影响因子:
10.7
通讯作者:
MCPEAK, H
中科院分区:
文献类型:
--
作者:
WILLIAMS, EM;BURROUGH, SLM;MCPEAK, H
The ability of the Transit-time Ultrasonic Breath Analyser (TUBA, GHG Medical Electronics GMBH, Zurich, Switzerland) to measure peak flow and tidal volume in the laboratory was tested using a variety of flow and pressure conditions, chosen to simulate the respiratory patterns of patients receiving mechanical ventilatory support. A stable zero baseline was achieved by acoustic damping of the TUBA flow sensor head. A piston pump was used to generate sinusoidal flow pattern, with a peak flow range from 0.1 to 1.51.s(-1). The calculated peak flow matched the peak flow measured by the TUBA. The TUBA accurately measured tidal volumes (+/- 10%) delivered using three different flow patterns over a range of volumes from 0.25 to 11. We conclude, that once modified, the TUBA can provide an accurate measurement of peak flow and tidal volume over a range of values likely to be encountered during mechanical ventilation of the lungs.